~/writing / essay

The Vendor Could Not Reproduce the Issue

On vendor accountability when the system behaves badly only at the customer site.

A hospital physicist reports a patient injury. The vendor’s field team flies out, runs the diagnostic suite, finds nothing, and files a report concluding the issue could not be reproduced. The physicist is told the machine is operating within specification.

A second injury at a different hospital. The vendor’s internal investigation points at a hardware component and ships a replacement part. After one of the early incidents, the vendor sends a two-page letter to the affected hospital explaining why the overdose the physicians are describing is not possible on this machine. Both machine failure and operator error, the letter says, can be ruled out.

A third injury. A fourth. A fifth. When the physicist at the third affected hospital asks the vendor whether there have been any other reports of overexposure, the vendor’s representatives, including the quality assurance manager, tell him that they know of no other accidents involving radiation overexposure on this machine. By the time of that conversation, the vendor is aware of two prior incidents and a lawsuit that has been pending for four months.

This is the Therac-25 investigation. It ran from 1985 to 1987. Six patients received massive radiation overdoses on a computer-controlled radiation therapy machine. Three died. The accidents involved multiple software defects, including a race condition in the operator interface and a separate setup-check failure involving a one-byte counter, in a machine whose previous model had hardware interlocks that had been quietly catching some of the same software problems for years. The interlocks were removed in the new model because the software was trusted to handle it.

I am not writing about Therac-25. Leveson and Turner’s case study has been required reading in safety engineering for thirty years and the technical lessons are well-rehearsed. I am writing about the part of the story that has not aged, which is the correspondence.

The pattern is familiar. A researcher reports an issue. The vendor cannot reproduce. The vendor’s internal review concludes the issue is not exploitable, or not severe, or not theirs. The researcher provides more detail. The vendor acknowledges the issue exists but disputes the impact. The researcher publishes. The vendor notes that no customers have been affected. A customer is affected. The vendor notes the affected customer was misconfigured.

Read that paragraph as describing AECL in 1986. Read it as describing any major vendor disclosure thread you saw last quarter. Both readings work. That is the point.

The mechanism in 1986 was not malice, or at least not only malice. The hospital physicist who eventually cracked the bug was Fritz Hager at East Texas Cancer Center in Tyler. After two patients at his facility were injured by doses he had been told the machine could not deliver, he sat at the console with the operator who had been there during both accidents. They typed prescription after prescription, late on a Friday, into the weekend, trying to reproduce Malfunction 54. Eventually they did. The trick was speed. If the operator entered the data fast enough, the machine bypassed the interlock check and fired the high-current beam without the target in place. Hager taught AECL how to reproduce it. AECL had spent months telling hospitals it could not be reproduced because their engineers were not typing fast enough to trigger it. The operators at the affected hospitals were typing faster, because they used the machine every day.

The vendor’s investigation methodology was not designed to find what their customers were doing. It was designed to confirm what the vendor expected to find.

The same mechanism is operating now. A vendor’s internal reproduction environment looks like a vendor’s internal reproduction environment. It does not look like a customer environment that has been running the product for three years with the configurations real customers actually use. When the vendor says they cannot reproduce, they often mean they cannot reproduce in the environment where they looked. This is not the same statement.

There is a second thread worth pulling on, which is the error message.

The Therac-25 was not silent during the overdoses. It threw an error code: Malfunction 54. The user manual did not document what Malfunction 54 meant. A technician later testified, after the deaths, that the closest manual entry described it as “dose input 2,” which meant the radiation delivered was either too high or too low. Operators were trained to press P to proceed past pause-class errors because pause-class errors fired constantly and almost always meant nothing. The fatal overdoses produced exactly the same error as the routine ones. The signal was there. The signal was indistinguishable from noise.

Anyone who has spent a week in a SOC understands it intuitively. The alert fires. It has fired forty times this month and has never meant anything. The analyst clicks through. Somewhere in the volume, the signal that mattered fired with the same color and the same priority as the signal that did not. The legible record will show the alert was generated. The forensic question after the breach is always whether the alert was generated. The alert is almost always generated. The alert is Malfunction 54.

The third thread is the interlocks.

The Therac-20 had hardware fuses that would blow if the machine entered an unsafe state. The Therac-25 inherited software from the Therac-20 and removed the fuses. The software had been buggy on the Therac-20 too. Nobody knew, because the fuses had been blowing instead of patients getting hurt, and a blown fuse is not a bug report. AECL believed the software was safe because the field history was clean. The field history was clean because the hardware was catching what the software missed.

This is the story of every defense-in-depth control that gets removed because a newer control has made it redundant. The newer control is trusted because the older control’s history is clean. The older control’s history is clean because it has been working. When you remove it, you discover what it was doing.

I am not going to provide a list of modern parallels here because the list would be the wrong artifact. The exercise the reader should do is to look at their own environment and find the controls that have been removed or downgraded in the last five years on the argument that something newer made them unnecessary. Network segmentation that got thinned because the identity layer is strong. MFA prompts that got reduced because the device posture signal is good. Endpoint controls that got loosened because the network controls are tight. Each of those decisions might be correct. The Therac-25 question is whether anyone has tested the surviving controls in isolation, with the removed ones genuinely gone, against the workloads that are actually running.

The Therac-25 case is taught as a story about safety-critical software. It is also a story about how organizations respond to evidence that contradicts what they have already concluded. AECL was not a careless company. It was a Crown corporation with a successful predecessor product and a competent engineering staff. It produced a machine that killed people, and then spent more than a year telling the people closest to the evidence that the evidence was wrong. The institutional response to the first injury report shaped the institutional response to every subsequent one.

The cybersecurity profession reads breach reports the way AECL read injury reports. We are looking for the misconfigured customer. We are looking for the user error. We are looking for the explanation that lets the product remain trusted.

The misconfigured customer is sometimes us.